The Regulator Exchange program:
$30.00 Flat rate for maintenance, parts extra with customer approval.
Independent Medical will send one loaner regulator for each regulator you have, to use while you send your regulator in for annual maintenance.
This includes:
· Visual inspection and cleaning of oxygen regulator. (Looking for damaged or missing parts)
· Leak test (External and Internal)
· Pressure regulation flow test. (The testing of the regulator for proper pressure regulation)
· Constant flow test. (The testing of liter flow settings)
· Gauge test. (Checking for calibration)
· CGA washer seal (Brass / Viton)
· Final physical inspection and repair. (When the approved work and repairs are completed)
· Documentation. (Maintenance record)
Upon completion of maintenance your regulators will be sent back with documentation, and easy to use box and label to return all loaner regulators back to Independent Medical.
|